🇨🇦 Olympus HF Cable
Hazard/Reason: Olympus received a complaint involving a thermal event with an HF cable during a procedure, causing a user burn. Investigation showed the device was used beyond IFU-defined service life, leading to insulation wear, corrosion, and failure. A four-year review identified 109 complaints, with no serious injuries and 17 minor burns.
This page links to official sources. Verify model/lot/batch details in the official notice.