🇨🇦 Mixing Bowl And Spatula, 20- Pack
Hazard/Reason: Zimmer Biomet is conducting a lot specific medical device recall removal for the mixing bowl and spatula. Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Seal integrity testing conducted internally showed potential sterility breach in 7.5% of the tested units that had visual seal nonconformities. The devices are distributed in 20-pack boxes.
The defectiveness was discovered on November 19, 2025. Issue was identified through the complaints handling process.
This page links to official sources. Verify model/lot/batch details in the official notice.