🇨🇦 Azurion and Allura Xper Systems
Hazard/Reason: The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. these fumes may activate environmental detection systems.