🇨🇦 Azurion Systems

Recall date: May 7, 2026Country: CACategory: medical_devices
Hazard/Reason: Philips has identified that the float tabletop control on the control module of Azurion systems may experience mechanical wear over time. The float tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the float tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges.