🇨🇦 CareLink™ SmartSync™ Device Manager
Hazard/Reason: A complaint from the CareLink network reported a sudden decrease in the Remaining Longevity of a patient's Micra VR CIED from 3 years to 6 months following a remote transmission, which was later found to be incorrect. Analysis revealed a calculation issue in SmartSync and CareLink for Micra VR2 and Micra AV2 devices, occurring under specific conditions. This miscalculation stems from a software error where a function received an integer instead of a real data type, defaulting the average pacing rate to 60 BPM and resulting in inaccurate Remaining Longevity estimates.
This page links to official sources. Verify model/lot/batch details in the official notice.