SafeRecalls
Back to search

🇨🇦 FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Recall date: October 6, 2026Country: CACategory: medical_devices
Hazard/Reason: A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
This page links to official sources. Verify model/lot/batch details in the official notice.